Operator’s Manual DXC100A Differential Probe Pair
DXC100A Differential Probe Pair 8 922259-00 Rev A 1. DA1855A and Oscilloscope Setup Follow the sequence shown in the previous Short Calibration Proc
Operator’s Manual 922259-00 Rev A 9 DA1855A has no probe coding input, and the wire is not used. Skip to Step 2. It is important to use the right
DXC100A Differential Probe Pair 10 922259-00 Rev A b. Adjust R6 (-X10 DC) to bring the DA1855A output to center screen (0 volts) on the oscilloscope
Operator’s Manual 922259-00 Rev A 11 6. +Input X10 LF Compensation NOTE: Low frequency compensation of the DXC100A probe attached to the +Input is d
DXC100A Differential Probe Pair 12 922259-00 Rev A 8. +Input X10 HF Transient Response a. Press the –Input Off button. b. Set the BW Limit to Ful
Operator’s Manual 922259-00 Rev A 13 10. +Input X100 LF Compensation NOTE: Low frequency compensation of the DXC100A probe attached to the +Input is
DXC100A Differential Probe Pair 14 922259-00 Rev A 11. X100 LF CMRR a. Continuing from the previous step, disconnect the probe from the pulse gener
Operator’s Manual 922259-00 Rev A 15 d. Set the oscilloscope to 5mV/div. e. Adjust R18 (-X100 HF1) for minimum displayed signal. The residual disp
DXC100A Differential Probe Pair 16 922259-00 Rev A When the +probe has been properly adjusted it should be used as a reference. CMRR adjustments made
Operator’s Manual 922259-00 Rev A 17 Figure 2, The DXC100A Schematic
DXC100A Differential Probe Pair 18 922259-00 Rev A Certifications This section contains the instrument’s Electromagnetic Compatibility (EMC), Safety
Operator’s Manual 922259-00 Rev A 19 Australia & New Zealand Declaration of Conformity—EMC Probe complies with the EMC provision of the Radio Com
DXC100A Differential Probe Pair 20 922259-00 Rev A Contact Teledyne LeCroy Teledyne LeCroy Service Centers United States and Canada - World Wide Cor
© 2013 Teledyne LeCroy, Inc. All rights reserved. Unauthorized duplication of Teledyne LeCroy documentation materials other than for internal sale
DXC100A Differential Probe Pair 2 922259-00 Rev A Safety Instructions This section contains instructions that must be observed to keep this oscillosc
Operator’s Manual 922259-00 Rev A 3 Keep product surfaces clean and dry. Do not submerge probe. Clean with a water- or alcohol-moistened cloth. Do no
DXC100A Differential Probe Pair 4 922259-00 Rev A A Word about Differential Amplifiers and Probes When using a differential amplifier it is very imp
Operator’s Manual 922259-00 Rev A 5 Probe Grounding The DXC100A Probe Pair is supplied with accessories allowing for three probe ground connection m
DXC100A Differential Probe Pair 6 922259-00 Rev A Once you are accustomed to making these adjustments, they become quick and easy. Part 2 can be atte
Operator’s Manual 922259-00 Rev A 7 Full Calibration Procedure If the probe pair has been repaired or is to be used on a DA1855A other than the one
Comments to this Manuals