Instruction Manual CP150 Current Probe
CP150 Current Probe 4 922175-00 Rev A Operation Precautions The sensor head is a precision assembly consisting of a molded component with a ferri
Instruction Manual 922175-00 Rev A 5 Operation with a Teledyne LeCroy Oscilloscope The probe can be controlled through the Vertical Setup dialog fo
CP150 Current Probe 6 922175-00 Rev A Care and Maintenance Cleaning The exterior of the probe and cable should be cleaned only using a soft cloth m
Instruction Manual 922175-00 Rev A 7 Returning a Defective Probe Contact your local Teledyne Lecroy sales representative to find out where to return
CP150 Current Probe 8 922175-00 Rev A Performance Verification This procedure can be used to verify the warranted characteristics of the CP150 Curr
Instruction Manual 922175-00 Rev A 9 Description Minimum Requirements Test Equipment Examples Function Generator 50 Hz sine wave output 3 Vrms
CP150 Current Probe 10 922175-00 Rev A Check LF Accuracy Figure 1. LF Accuracy Test set up. 1. Set the Function generator to 50 Hz sine wave. O
Instruction Manual 922175-00 Rev A 11 Adjustment Procedure This procedure can be used to adjust the warranted characteristics of the CP150 Current Pr
CP150 Current Probe 12 922175-00 Rev A Description Minimum Requirements Test Equipment Examples Patch Cables (4 required) Male Banana to Male B
Instruction Manual 922175-00 Rev A 13 1. Set the Function Generator to 50 Hz, sinewave output at 1.7 Vrms with 50 Ω output. 2. Connect BNC male of
CP150 Current Probe 14 922175-00 Rev A Specifications Nominal Characteristics Nominal characteristics describe parameters and attributes which are
Instruction Manual 922175-00 Rev A 15 Environmental Characteristics Operating Temperature Humidity 0 to 40 °C (32 °F to 104 °F) ≤ 80% relative hum
CP150 Current Probe 16 922175-00 Rev A Graphs Figure 3. Maximum input current vs. frequency. Figure 4. Insertion impedance vs. frequency.
Instruction Manual 922175-00 Rev A 17 Certifications This section contains the instrument’s Electromagnetic Compatibility (EMC), Safety and Environme
CP150 Current Probe 18 922175-00 Rev A CISPR 11:2003 Radiated and Conducted Emissions, Group 1, Class A, in accordance with EN61326-1:2006 and EN61
Instruction Manual 922175-00 Rev A 19 Contact Teledyne LeCroy Teledyne LeCroy Service Centers United States and Canada - World Wide Corporate Offic
CP150 Current Probe 20 922175-00 Rev A Appendix A Performance Verification Test Record Photocopy this page and record the results of measurements m
CP150 Current Probe Instruction Manual January 2013
700 Chestnut Ridge Road Chestnut Ridge, NY, 10977-6499 Tel: (845) 425-2000 Fax: (845) 578 5985 teledynelecroy.com © 2013 Teledyne LeCroy, Inc.
Instruction Manual 922175-00 Rev A i TABLE OF CONTENTS Safety Instructions ...
CP150 Current Probe ii 922175-00 Rev A Preliminary Procedure...
Instruction Manual 922175-00 Rev A 1 Safety Instructions This section contains instructions that must be observed to keep this oscilloscope accessory
CP150 Current Probe 2 922175-00 Rev A Be careful not to damage the insulation surface when making measurements. Never install or remove the p
Instruction Manual 922175-00 Rev A 3 Overview Description The CP150 uses a combination of Hall effect and transformer technology which enables measur