Operator’s Manual ZS1000, ZS1500, ZS2500 High-Impedance, Active Probe
ZS Series High-Impedance, Active Probes Features and Accessories The ZS1000, ZS1500, and ZS2500s probes are provided with numerous features and access
Operator’s Manual Tips Straight Tip The straight tip is rugged and designed for general probing. Fits in either probe socket. (PK-ZS-001) IC Lead T
ZS Series High-Impedance, Active Probes Tips, continued Discrete SMD Tip The crescent shape of this tip is designed to fit tightly on capacitors, res
Operator’s Manual Grounds Offset Ground The offset pin is designed to be attached to either socket of the probe head. The offset pin is the highest
ZS Series High-Impedance, Active Probes Leads While longer leads provide greater flexibility when connecting the probe to a circuit, the added inducta
Operator’s Manual Probe Holder Freehand Probe Holder The FreeHand lets you focus on the oscilloscope screen instead of on maintaining contact to mult
ZS Series High-Impedance, Active Probes Operation with a Teledyne LeCroy Oscilloscope When the ZS1000, ZS1500, or ZS2500 probe is connected to any com
Operator’s Manual High Frequency Measurements Probe Input Loading When you touch a probe to the circuit under test, the probe will affect your measure
ZS Series High-Impedance, Active Probes The resonant frequency of a series LC circuit can be raised by decreasing the inductance, capacitance or both.
Operator’s Manual Where Ctotal is the combined probe and circuit capacitance and Rtotal is combined circuit and probe resistance. For a setup where Ct
ZS Series High-Impedance, Active Probes Probe Impedance and Equivalent Circuit ZS1x00 Impedance ZS1x00 Equivalent Circuit 10100100010000100000100000
Operator’s Manual ZS2500 Impedance ZS2500 Equivalent Circuit 1101001000100001000001000000100Hz 1kHz 10kHz 100kHz 1MHz 10MHz 100MHz1000MHz10000MHz 92
ZS Series High-Impedance, Active Probes Care and Maintenance Cleaning The exterior of the probe and cable should be cleaned, using a soft cloth moiste
Operator’s Manual Return shipment should be prepaid. Teledyne LeCroy cannot accept COD or Collect Return shipments. We recommend air-freighting. Foll
ZS Series High-Impedance, Active Probes Replacement Parts The probe connection accessories and other common parts can be ordered through the North Ame
Operator’s Manual Because the input and output connector types may vary on different brands and models of test instruments, additional adapters or cab
ZS Series High-Impedance, Active Probes 3. Turn on the other test equipment and allow them to warm up for the manufacturer’s recommended timeframe. 4
Operator’s Manual Verification Procedure A. Output Zero Voltage Output Zero Voltage Test Setup 1. Connect one end of a BNC cable to the female BNC c
ZS Series High-Impedance, Active Probes B. LF Attenuation Accuracy LF Attenuation Accuracy Setup 1. Disconnect the BNC tee at the power supply from
Operator’s Manual 6. Set the mode of the function generator to sine wave, the frequency to 70 Hz and the output amplitude to 5 Vrms ±10 mV as measure
ZS1000, ZS1500, ZS2500 High-Impedance Active Probes Operator’s Manual
ZS Series High-Impedance, Active Probes 20. Remove the banana plug adapter from the DMM and connect the precision 50 Ω terminator to the DMM, making s
Operator’s Manual Item Tested Item Serial Number Date Technician ZS Equipment Used Instrument Model Serial Number Calibration Due Date Osc
ZS Series High-Impedance, Active Probes Reference Material Specifications Please refer to the Teledyne LeCroy website at teledynelecroy.com for detail
Operator’s Manual Safety Compliance EC DECLARATION OF CONFORMITY – LOW VOLTAGE The probe meets intent of EC Directive 2006/95/EC for Product Safety. C
ZS Series High-Impedance, Active Probes Contact Teledyne LeCroy Teledyne LeCroy Service Centers United States and Canada - World Wide Corporate Offic
© 2014 Teledyne LeCroy, Inc. All rights reserved. Unauthorized duplication of Teledyne LeCroy documentation materials other than for internal sale
Operator’s Manual TABLE OF CONTENTS Safety Instructions ... 1 Symbols...
ZS Series High-Impedance, Active Probes Care and Maintenance ... 16 Cleaning ..
Operator’s Manual Safety Instructions This section contains instructions that must be observed to keep this oscilloscope accessory operating in a corr
ZS Series High-Impedance, Active Probes Operating Environment The accessory is intended for indoor use and should be operated in a clean, dry environ
Operator’s Manual Standard Accessories The ZS1000, ZS1500, and ZS2500 probes are shipped with the following standard accessories: Standard Accessory